GuideSheath Kit 2 Quick Reference GuideSingle Use Guide Sheath Kit
K-401/402/403/404
Below are important steps for handling the GuideSheath Kit 2. For detailed operating instructions, be sure to follow the instruction manual included in the device package.
1 Types of Stoppers
The GuideSheath Kit 2 comes with two types of stoppers – one for an ultrasonic probe and one for EndoTherapy accessories.2 Operation of ET/US Stoppers
<Preparation of US Stopper>
Insert the ultrasonic probe into the guide sheath’s accessory port so that the probe tip protrudes slightly from the guide sheath tip. Attach the US stopper to the ultrasonic probe by inserting the probe to the end of the slit.
<Preparation of ET Stopper>
Insert the EndoTherapy device into the guide sheath’s accessory port so that the device tip protrudes slightly from the guide sheath tip. Attach the ET stopper to the EndoTherapy device by inserting the EndoTherapy device to the end of the slit.
3 Attachment of Stopper to the Guide Sheath
-
Do not insert the guide sheath into the endoscope when it is not combined with any EndoTherapy accessory or ultrasonic probe. Doing so could result in bending or breaking the guide sheath.
-
When inserting the guide sheath, hold it close to the biopsy valve and keep it as straight as possible, then insert it gently into the endoscope.
Caution
Specifications, design and accessories are subject to change without any notice or obligation on the part of the manufacturer.